Attempt of absolute analysis with spark source mass spectrometry
Creators
- 1. Commissariat a l'Energie Atomique, Grenoble (France). Centre d'Etudes Nucleaires
Description
By means of a graphical method developed in our laboratory, we have studied the linearity of the response of the MS-7 mass spectrometer for impurity determinations over a concentration range of 1 to 1000 ppm (parts per million of atoms). This method consist in transforming optical density measurements into 'true intensities', which are plotted on a logarithm-scale paper against the exposures. A moving transparent ruler graduated at the inverse scale of the exposures allows us to determine directly on the graph, the concentration of impurities in ppm. We have used this method for the determination of sensitivity coefficients in standard samples such as Al, Fe, Cu, Ni, Zr, and non conducting powders like SiO2 and Al2O3. This study shows that, for the samples studied, the sensitivity coefficients are practically independent of the matrix and the concentration. Moreover the results show the possibility of obtaining an absolute analysis by spark source mass spectrometry. (author)
Availability note (English)
Available from INIS in electronic formAbstract (French)
Au moyen d'une methode graphique developpee ou laboratoire, nous avons etudie la linearite des determinations d'impuretes au spectrometre de masse, type MS-7, dans un domaine de concentration allont de 1 a 1000 ppm. Cette methode consiste en la transformation des mesures de densite optique en 'intensite vraie'. Ces dernieres mesures sont portees sur un graphique logarithmique et une reglette transparente graduee a l'inverse de l'echelle des expositions nous permet de determiner directement la teneur des impuretes en ppm (partie par million d'atomes). Nous avons utilise cette methode pour la determination des coefficients de sensibilite dans des echantillons etalons tels que Al, Fe, Cu, Ni, Zr, et des poudres non conductrices genre SiO2, Al2O3. Cette etude a montre que, pour les echantillons etudies, les coefficients de sensibilite sont pratiquement independants de la matrice et de la concentration. De plus, les resultats obtenus montrent la possibilite d'effectuer une analyse absolue par spectrometrie de masseFiles
36115111.pdf
Files
(765.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:eceb3878663e50ab67dba7fb7617cd76
|
765.6 kB | Preview Download |
Additional details
Additional titles
- Original title (French)
- Essai d'analyse absolue par spectrometrie de masse a etincelles
Publishing Information
- Imprint Pagination
- 40 p.
- Report number
- CEA-R--2721
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 36115111
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Thesis
- Descriptors DEI
- CALIBRATION; IMPURITIES; QUANTITATIVE CHEMICAL ANALYSIS; SENSITIVITY; SPARK MASS SPECTROMETERS
- Descriptors DEC
- CHEMICAL ANALYSIS; MASS SPECTROMETERS; MEASURING INSTRUMENTS; SPECTROMETERS
Optional Information
- Notes
- 22 refs.