Published November 15, 1986
| Version v1
Journal article
Chevron of microchannel plates as position-sensitive detector for β-spectrometers
- 1. AN Kazakhskoj SSR, Alma-Ata. Inst. Yadernoj Fiziki
Description
The detection efficiency and position resolution of a microchannel plate chevron for electron beams of 12 to 300 keV energy at 00 to 450 angles of incidence have been investigated. The chevron is shown to ensure good stability, high detection efficiency and high position resolution in the energy range under study. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 252
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 83-86
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 18019833
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA DETECTION; BETA SPECTROMETERS; EFFICIENCY; ELECTRON CAPTURE DECAY; ELECTRON SPECTRA; K CAPTURE; L CAPTURE; M CAPTURE; MICROCHANNEL ELECTRON MULTIPLI; POSITION SENSITIVE DETECTORS; SPATIAL RESOLUTION; STABILITY; THULIUM 169
- Descriptors DEC
- BETA DECAY; CHARGED PARTICLE DETECTION; DECAY; DETECTION; ELECTRON MULTIPLIERS; ELECTRON TUBES; INTERMEDIATE MASS NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NUCLEAR DECAY; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTION; RADIATION DETECTORS; RARE EARTH NUCLEI; RESOLUTION; SPECTRA; SPECTROMETERS; STABLE ISOTOPES; THULIUM ISOTOPES