Published May 1, 2018 | Version v1
Journal article

Performance studies of X3 silicon detectors for the future ELISSA array at ELI-NP

  • 1. Extreme Light Infrastructure-Nuclear Physics, Horia Hulubei National Institute for R and D in Physics and Nuclear Engineering, Bucharest-Magurele, 077125 (Romania)

Description

ELISSA is an array of silicon strip detectors under construction at the ELI-NP facility for measurements of photodissociation reactions using high-brilliance, quasi monoenergetic gamma beams. The detection system consists of 35 single-sided position-sensitive X3 detectors arranged in a cylindrical configuration and eight QQQ3 detectors as end-caps. A batch of forty X3 detectors have been tested at ELI-NP. The energy and position resolution, ballistic deficit, leakage currents, and depletion voltage were measured and analyzed. Measurements of the energy resolution were carried out using two read-out electronic chains, one based on multichannel preamplifiers and another based on multiplexers.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/13/05/T05006

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
13
Journal Issue
05
Journal Page Range
p. T05006
ISSN
1748-0221