Published June 21, 2013 | Version v1
Journal article

TOF PET offset calibration from clinical data

  • 1. Department of Radiology, The University of Pennsylvania, Philadelphia, PA 19104 (United States)

Description

In this paper, we present a timing calibration technique for time-of-flight positron emission tomography (TOF PET) that eliminates the need for a specialized data acquisition. By eliminating the acquisition, the process becomes fully automated, and can be performed with any clinical data set and whenever computing resources are available. It also can be applied retroactively to datasets for which a TOF offset calibration is missing or suboptimal. Since the method can use an arbitrary data set to perform a calibration prior to a TOF reconstruction, possibly of the same data set, one also can view this as reconstruction from uncalibrated data. We present a performance comparison with existing calibration techniques. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/58/12/4031

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
58
Journal Issue
12
Journal Page Range
p. 4031-4046
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44114586
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CALIBRATION; DATA ACQUISITION; DATASETS; POSITRON COMPUTED TOMOGRAPHY; TIME-OF-FLIGHT METHOD
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOCUMENT TYPES; EMISSION COMPUTED TOMOGRAPHY; TOMOGRAPHY