NaI(Tl) scintillator read out with SiPM array for gamma spectrometer
Creators
Description
The NaI(Tl) scintillator is widely used in gamma spectrometer with photomultiplier tube (PMT) readout. Recently developed silicon photomultiplier (SiPM) offers gain and efficiency similar to those of PMT, but with merits such as low bias voltage, compact volume, low cost, high ruggedness and magnetic resonance compatibility. In this study, 2-in. and 1-in. NaI(Tl) scintillators were readout with SiPM arrays, which were made by tiling multiple SiPMs each with an active area of 6×6 mm2 on a printed circuit board. The energy resolutions for 661.6 keV gamma rays, obtained with Φ2×2 in. scintillator coupled to 6×6 ch SiPM array and Φ1×1 in. scintillator coupled to 4×4 ch SiPM array were 7.6% and 7.8%, respectively, and were very close to the results obtained with traditional bialkali PMT (7.3% and 7.6%, respectively). Scintillator coupled to photodetector with smaller area was also studied by adding a light guide or using scintillator with tapered head. The latter showed better performance than using light guide. The 1-in. NaI(Tl) scintillator with tapered head coupled to 2×2 ch SiPM array achieved 7.7% energy resolution at 661.6 keV, the same as that obtained with standard Φ1×1 in. scintillator coupled to 4×4 ch SiPM array. While the 2-in. scintillator with similar geometry showed degraded energy resolution, 10.2% at 661.6 keV, but could still be used when high efficiency is preferred over energy resolution.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2017.01.068Additional details
Identifiers
- DOI
- 10.1016/j.nima.2017.01.068;
- PII
- S0168-9002(17)30157-2;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 851
- Journal Page Range
- p. 118-124
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48082788
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- EFFICIENCY; ELECTRIC POTENTIAL; ENERGY RESOLUTION; GAMMA RADIATION; GAMMA SPECTROMETERS; GAMMA SPECTROSCOPY; KEV RANGE 100-1000; MAGNETIC RESONANCE; NAI DETECTORS; OPTICAL FIBERS; PHOTODETECTORS; PHOTOMULTIPLIERS; PRINTED CIRCUITS; READOUT SYSTEMS; SILICON; SODIUM IODIDES
- Descriptors DEC
- ALKALI METAL COMPOUNDS; ELECTROMAGNETIC RADIATION; ELECTRONIC CIRCUITS; ELEMENTS; ENERGY RANGE; FIBERS; HALIDES; HALOGEN COMPOUNDS; INORGANIC PHOSPHORS; IODIDES; IODINE COMPOUNDS; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; PHOSPHORS; PHOTOTUBES; RADIATION DETECTORS; RADIATIONS; RESOLUTION; RESONANCE; SCINTILLATION COUNTERS; SEMIMETALS; SODIUM COMPOUNDS; SODIUM HALIDES; SOLID SCINTILLATION DETECTORS; SPECTROMETERS; SPECTROSCOPY
Optional Information
- Copyright
- Copyright (c) 2017 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.