Published January 1995 | Version v1
Journal article

Production, administration and disposal of cyclotron produced shortlived radioactive gases for positron emission tomography studies at the Austin Repatriation Medical Centre, Melbourne

  • 1. Austin Hospital, Heidelberg, VIC (Australia)
  • 2. Melbourne Univ., Parkville, VIC (Australia). School of Physics

Description

Positron Emission Tomography (PET) Centre is operational at the Austin Repatriation Medical Centre, Melbourne. The major equipment consists of a 10 MeV cyclotron and a whole body PET scanner. Radioactive gases produced and used directly in clinical studies include [15O]O2, [15O]CO, and [15O]CO2, whilst [11C]CO2 is also produced for use in radiochemistry syntheses. Radioactivity delivery rates of 3.7, 3.3, and 1.6 GBq/min to the scanner suite have been achieved for [15O]O2, [15O]CO2, and [15O]CO respectively, and batch productions of 36.3 GBq of [11C]CO2 have been produced. The production. patient administration and disposal of the short-lived radioactive gases has been achieved in compliance with radiation protection principles. Radioactive gas doses of 1.7 GBq are administered to patients with less than 0.02 MBq/m3 leakage into the scanner suite. Less than 13 MBq of [ 15O]-labelled gases are released into the environment per patient study at a concentration of 0.018 MBq/m3. Annually less than 2 GBq is expected to be released into the environment. The centre design and first four months' experience of radioactive gas production, administration and disposal is presented. 5 refs., 4 tab., 1 fig

Additional details

Publishing Information

Journal Title
Radiation Protection in Australia
Journal Volume
13
Journal Issue
1
Journal Page Range
p. 26-30.
ISSN
0729-7963
CODEN
RPAUDH

Conference

Title
18. annual conference of the Australian Radiation Protection Society.
Dates
Oct 1993.
Place
Sydney (Australia).