Application of automatic tube current modulation technology in PET/CT
Creators
- 1. Department of Nuclear Medicine, Chongqing University Cancer Hospital, Chongqing Cancer Institute, Chongqing Cancer Hospital, Chongqing (China)
Description
Objective: To investigate the effects of automatic tube current modulation (ATCM) on CT image quality and effective dose (ED) of PET/CT of 90 patients. Methods: PET/CT of 90 patients from October 2017 to July 2018 were randomly divided into three groups, A, B, and C by systematic sampling (30 cases in each group), and CT was scanned by ATCM. The tube current interval of group A and B was 60-240 mA, and the noise index (NI) was 10 and 15;the tube current interval of group C was 60-180 mA, and the NI was 15. The CT volume dose index (CTDIvol) and dose length product (DLP) were recorded, and the ED was estimated according to the formula. The CT image quality of neck, chest, abdomen, and pelvis were evaluated blindly by two and above attending physicians in nuclear medicine to calculate signal:to-noise ratio (SNR) by measuring CT value and noise value of the CT image. Noise value and SNR were compared with variance analysis. CTDIvol, DLP, and ED were compared with Kruskal-Wallis and with Nemenyi in pairs. Results: Differences in noise value [(11.90 ± 2.83)-(26.03 ± 3.74)] and SNR [(2.03 ± 0.34)-(4.35 ± 0.71)] among three groups were statistically significant (F = 38.01-64.20, F = 32.09-81.62, and all P < 0.05), and the CT image quality met the clinical diagnostic requirements. CTDIvol were (12.44 ± 0.53), (9.39 ± 2.01), and (7.05 ± 1.03) mGy, DLP were (998.45 ± 96.04), (741.60 ± 168.87), and (571.29 ± 97.41) mGy · cm, and ED were (14.98 ± 1.44), (11.12 ± 2.53), and (8.57 ± 1.46) mSv among three groups, with statistically significant differences (x2 = 62.18, 57.19, 57.16, and all P < 0.05). Group C was lower than group A (x2 = 56.55, P < 0.05) with statistically significant differences. Conclusion: ATCM technology can ensure the image quality while reducing the ED effectively by modulating the current interval and NI reasonably. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- International Journal of Radiation Medicine and Nuclear Medicine
- Journal Volume
- 43
- Journal Issue
- 1
- Journal Page Range
- p. 22-26
- ISSN
- 1673-4114
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 54049938
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABDOMEN; CHEST; COMPARATIVE EVALUATIONS; EFFECTIVE RADIATION DOSES; ELECTRIC CURRENTS; IMAGES; MODULATION; NECK; PELVIS; POSITRON COMPUTED TOMOGRAPHY; QUALITY CONTROL; SIGNAL-TO-NOISE RATIO; TUBES
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; CONTROL; CURRENTS; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DOSES; EMISSION COMPUTED TOMOGRAPHY; EVALUATION; RADIATION DOSES; TOMOGRAPHY
Optional Information
- Notes
- 4 tabs., 16 refs.; http://dx.doi.org/10.3760/cma.j.issn.1673-4114.2019.01.005