The robustness of prostate radiotherapy for patients with hip prosthesis
Creators
- 1. Department of Medical Physics, Tampere University Hospital, 33521 Tampere (Finland)
Description
The aim of this study was to investigate prostate radiotherapy techniques for the patients with hip prosthesis in 4 different field setups. Volumetric Modulated Arc Therapy (VMAT) technique was used in 4 different cases: (1) using full VMAT arcs (VMATF); (2) same arcs as in case 1 but with avoidance sectors (VMATASEC); (3) as case 2 but with the addition of a lateral static field through the prosthesis (VMATASEC+STAT); (4) as in case 1 but with an automated structure avoidance option to avoid irradiation through the prosthesis (VMATASTR). Fifteen previously treated prostate patients were retrospectively selected to this study. Treatment plans were created for all patients using all 4 techniques. The potential prosthesis misalignment in the treatment setup was modeled by moving the prosthesis 0.5, 1.0, and 1.5 cm ventrally and dorsally and recalculating the plans in each case. For VMATASEC, the dose parameters for organs at risk were the highest and the dose coverage of the target volume was the poorest when compared to the other techniques. For VMATASEC+STAT, the movement of the prosthesis changed the target dose distribution the most. VMATF and VMATASTR fulfilled the planning criteria the best, even when the prosthesis was misaligned. VMATF radiated through the prosthesis more than VMATASTR and increased the dose near the prosthesis surface when compared to VMATASTR. VMATASTR and VMATF were the most robust techniques for the patients with the hip prosthesis considering plan quality and the effect of positioning errors. The increased prosthesis surface dose with VMATF and possible dose calculation uncertainties favors the use of VMATASTR.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.meddos.2020.11.007Additional details
Identifiers
- DOI
- 10.1016/j.meddos.2020.11.007;
- PII
- S0958394720301631;
Publishing Information
- Journal Title
- Medical Dosimetry
- Journal Volume
- 46
- Journal Issue
- 3
- Journal Page Range
- p. 212-218
- ISSN
- 0958-3947
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54003112
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DOSIMETRY; ERRORS; HEALTH HAZARDS; IRRADIATION; PATIENTS; PLANNING; POSITIONING; PROSTATE; PROSTHESES; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- BODY; DOSES; GLANDS; HAZARDS; MALE GENITALS; MEDICAL SUPPLIES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY
Optional Information
- Copyright
- Copyright (c) 2020 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.