Published July 21, 1996 | Version v1
Journal article

Design of a divergence and alignment indicator

  • 1. Virginia Univ., Charlottesville, VA (United States). Dept. of Mechanical, Aerospace and Nuclear Engineering

Description

The divergence and alignment indicator (DAI) is an extension of the ASTM E803 L/D thermal neutron radiography L/D device that allows the user to determine both the beam centerline and the beam divergence. The DAI was made using aluminium plate and rods, and incorporated cadmium wire for contrast. Circular symmetry was utilized to simplify manufacture. The DAI was placed with the five posts against the film cassette or radioscopic imaging device in the physical center of the beam. The DAI was perpendicular to the selected beam radius when the front and back center Cd wire images overlap. The degree of misalignment was indicated by their image positions. After the DAI was aligned, analysis of the cadmium wire ''+'' image spacing yielded the beam divergence. The DAI was tested in a neutron beam which has an L/D of 30 but a small degree of divergence. The DAI was also imaged using an X-ray source. The point source predictions of Cd wire image locations showed good agreement with those measured from the X-ray radiograph. The neutron radiographic locations could be predicted using the point source equations, even though the neutron beam was a complex distributed source. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
377
Journal Issue
1
Journal Page Range
p. 58-62.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
2. international topical meeting on neutron radiography system design and characterization.
Dates
12-18 Nov 1995.
Place
Tokyo (Japan).