Comparison of absorbed dose of two protocols of tomographic scanning in PET/CT exams
Creators
- 1. Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Dept. de Engenharia Nuclear
- 2. Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Faculdade de Medicina
- 3. Centro Federal de Educação Tecnológica de Minas Gerais (CEFET-MG), Belo Horizonte, MG (Brazil)
Description
Positron Emission Tomography (PET) associated with Computed Tomography (CT) allows the fusion of functional and anatomical images. When compared to other diagnostic techniques, PET-CT subjects patients to higher levels of radiation, because two modalities are used in a single exam. In this study, the doses absorbed in 19 patient organs from the tomographic scan were evaluated. Radiochromic films were correctly positioned in the Alderson anthropomorphic simulator, male version. For evaluation, two whole body scan protocols were compared. For evaluation, two whole body scan protocols were compared. An increase of up to 600% in the absorbed dose in the pituitary was observed when the protocols were compared, with the lowest observed increase of approximately 160% for the liver. It is concluded that the dose from CT in patients submitted to PET-CT scanning is higher in the protocol used for diagnosis. Considering the high cost of PET-CT exam, in many cases it is preferable that the CT examination is of diagnostic quality, and not only for anatomical mapping, an argument based on the Principle of Justification
Files
49034213.pdf
Files
(410.1 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:351549e654f3e9a4842208f58216d8e7
|
410.1 kB | Preview Download |
Additional details
Additional titles
- Original title (Portuguese)
- Comparação da dose absorvida de dois protocolos de varredura tomográfica em exame de PET-CT
Publishing Information
- Imprint Pagination
- 3 p.
- Report number
- INIS-BR--20164
Conference
- Title
- Goiania 30 years later
- Acronym
- International joint conference RADIO 2017
- Dates
- 25-29 Sep 2017
- Place
- Goiania, GO (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 49034213
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ABSORBED RADIATION DOSES; COMPARATIVE EVALUATIONS; FILM DOSIMETRY; PHANTOMS; PITUITARY GLAND; POSITRON COMPUTED TOMOGRAPHY; WHOLE-BODY COUNTERS
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSES; DOSIMETRY; EMISSION COMPUTED TOMOGRAPHY; ENDOCRINE GLANDS; EVALUATION; GLANDS; MEASURING INSTRUMENTS; MOCKUP; ORGANS; RADIATION DETECTORS; RADIATION DOSES; STRUCTURAL MODELS; TOMOGRAPHY