Published June 1996 | Version v1
Journal article

Shielding design for a HiP-Ge detector system for the measurement of U and Th in environmental samples

  • 1. Universiti Teknologi Malaysia, Skudai, Johor Bahru (Malaysia). Dept. of Physics

Description

A low level high-purity-Ge detector system for the measurement of low level uranium and thorium has been set up in Nuclear Laboratory at Physics Department, UTM. A background reduction by a factor of 27 is achieved by materials selection, cryostat configuration and shielding. An analysis of the background components is given. The integrated count rate of 8.7 counts per second within the energy interval 50-1600 keV is achieved in this spectroscopy system

Additional details

Publishing Information

Journal Title
Jurnal Fizik UTM
Journal Volume
3
Journal Issue
1
Journal Page Range
p. 9-16.
ISSN
0128-8644