Purpose: There have been few reports for independent dose verification for Tomotherapy. We evaluated the accuracy and the effectiveness of an independent dose verification system for the Tomotherapy. Methods: Simple MU Analysis (SMU, Triangle Product, Ishikawa, Japan) was used as the independent verification system and the system implemented a Clarkson-based dose calculation algorithm using CT image dataset. For dose calculation in the SMU, the Tomotherapy machine-specific dosimetric parameters (TMR, Scp, OAR and MLC transmission factor) were registered as the machine beam data. Dose calculation was performed after Tomotherapy sinogram from DICOM-RT plan information was converted to the information for MU and MLC location at more segmented control points. The performance of the SMU was assessed by a point dose measurement in non-IMRT and IMRT plans (simple target and mock prostate plans). Subsequently, 30 patients' treatment plans for prostate were compared. Results: From the comparison, dose differences between the SMU and the measurement were within 3% for all cases in non-IMRT plans. In the IMRT plan for the simple target, the differences (Average±1SD) were −0.70±1.10% (SMU vs. TPS), −0.40±0.10% (measurement vs. TPS) and −1.20±1.00% (measurement vs. SMU), respectively. For the mock prostate, the differences were −0.40±0.60% (SMU vs. TPS), −0.50±0.90% (measurement vs. TPS) and −0.90±0.60% (measurement vs. SMU), respectively. For patients' plans, the difference was −0.50±2.10% (SMU vs. TPS). Conclusion: A Clarkson-based independent dose verification for the Tomotherapy can be clinically available as a secondary check with the similar tolerance level of AAPM Task group 114. This research is partially supported by Japan Agency for Medical Research and Development (AMED)

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
43
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48097539
Subject category
S60: APPLIED LIFE SCIENCES; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; PROSTATE; RADIATION DOSES; VERIFICATION
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSES; GLANDS; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RADIOTHERAPY; THERAPY; TOMOGRAPHY

Optional Information

Notes
(c) 2016 American Association of Physicists in Medicine