MTF analysis of the near-real time neutron radiography facility at MURR

  • 1. Missouri Univ., Columbia (USA). Research Reactor Facility

Description

Several neutron radiography systems designed to view transient processes on a real-time basis have been developed. With the advent of these different real-time systems comes the necessity to develop a means to quantitatively evaluate and compare these systems. A suitable method for measuring the resolution capabilities of the image-forming system is the determination of the modulation transfer function (MTF). The MTF is a measure of an imaging system's ability to reproduce the spatial frequencies present in an image. The system in use at the University of Missouri Research Reactor is described. (Auth.)

Additional details

Additional titles

Augmented title (English)
MTF = modulation transfer function

Publishing Information

Publisher
D. Reidel.
Imprint Place
Dordrecht (Netherlands)
ISBN
90-277-1528-9
Imprint Title
Neutron radiography
Imprint Pagination
1090 p.
Journal Page Range
p. 617-623.

Conference

Title
1. World conference on neutron radiography.
Dates
7-10 Dec 1981.
Place
San Diego, CA (USA).

Optional Information

Notes
Imprint:Includes author index.