Published November 2018 | Version v1
Journal article

Acceptance testing and commissioning of Radixact™ X9 tomotherapy unit: initial experience

  • 1. Department of Radiation Oncology, HCG Khubchandani Cancer Centre, Mumbai (India)

Description

Helical Tomotherapy is recognized as modern solution for radiotherapy to sculpture highly conformal dose to the target and minimal toxicity to OARs. A 6 MV linear accelerator is mounted on a ring gantry for image-guided intensity-modulated radiation therapy treatment. The major advantage of Radixact™ X9 system from its predecessor is that it offers increased dose rate of about 1000 cGy/min at isocenter and 10 rpm gantry rotation for faster image acquisition with reduced imaging dose rate. Like CT technology, gantry rotation and couch longitudinal translation are simultaneous that permits helical dose distribution up to 135 cm treatment length. A fan beam created by the collimator and jaws produces a maximum of 40 X 5 Cm2 field size at the isocenter. The linac can be operated at a lower voltage (3.5 MV) to produce CT images, which are used for patient image verifications. Radixact™ X9 tomotherapy unit was commissioned for clinical use at our center in June 2018

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
43
Journal Issue
suppl.1
Journal Page Range
p. 74-75
CODEN
JMPHFE

Conference

Title
39. annual conference of Association of Medical Physicists of India
Acronym
AMPICON 2018
Dates
2-4 Nov 2018
Place
Chennai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
50065605
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DOSE LIMITS; DOSIMETRY; LINEAR ACCELERATORS; RADIATION DOSES
Descriptors DEC
ACCELERATORS; DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; SAFETY STANDARDS; STANDARDS; THERAPY; TOMOGRAPHY