Development of a new integrated In-Vivo counting system at the QST
Creators
- 1. Center for Radiation Research and Education, Nagasaki University, 1-12-4 Sakamoto, Nagasaki 852-8523 (Japan)
- 2. National Institutes for Quantum Science and Technology, 4-9-1 Anagawa, Inage, Chiba 263-8555 (Japan)
Description
A new in-vivo counting system that functions as both a whole-body counter (WBC) and a lung counter (LC) was developed at the QST to enhance its dose assessment capability. This paper presents an overview of this system and the results of its performance tests. For use of the system as a WBC, three high purity germanium (HPGe) detectors installed in a 20-cm-thick iron shielding chamber are linearly arrayed over a subject lying on the bed, whereas two of the three HPGe detectors are placed over the subject's chest from side to side when using the system as an LC. The new in-vivo system was calibrated using three de-facto phantoms owned by the QST: an adult-male Bottle Manikin Absorption (BOMAB) phantom, a Lawrence Livermore National Laboratory (LLNL) phantom and a Japan Atomic Energy Research Institute (JAERI) phantom. Monte Carlo simulations were also performed to determine an optimum location for the three detector array in the WBC mode and revealed that the peak efficiency for the BOMAB phantom (662 keV) was little varied as long as the middle detector was placed above the thorax and abdomen parts of the phantom. The calculated peak efficiencies agreed well with the observed peak efficiencies for photons with energies over 100 keV. For lung counting, a tentative Minimum Detectable Activity of 241Am was evaluated as 9.5 Bq for a counting time of 30 minutes, and a Japanese male subject with an average chest wall thinness (2.27 cm). The developed system is now ready for use. (authors)
Additional details
Identifiers
- DOI
- 10.1093/rpd/ncac226;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 199
- Journal Issue
- 15-16
- Journal Page Range
- p. 1848-1852
- ISSN
- 0144-8420
Conference
- Title
- Joint International Conference on Individual Monitoring of Ionizing Radiation; NEUDOS-14: Neutron and Ion Dosimetry Symposium
- Acronym
- IM2022
- Dates
- 25-29 Apr 2022
- Place
- Krakow (Poland)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 55045344
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ABDOMEN; ABSORPTION; AMERICIUM 241; CHEST; COMPUTERIZED SIMULATION; GERMANIUM; HIGH-PURITY GE DETECTORS; IN VIVO; JAERI; LUNGS; MONTE CARLO METHOD; NUCLEAR ENERGY; PERFORMANCE; PHANTOMS; PHOTONS; RADIATION DOSES; SHIELDING; WHOLE-BODY COUNTERS
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; AMERICIUM ISOTOPES; BODY; BOSONS; CALCULATION METHODS; DOSES; ELEMENTARY PARTICLES; ELEMENTS; ENERGY; GE SEMICONDUCTOR DETECTORS; HEAVY NUCLEI; ISOTOPES; JAPANESE ORGANIZATIONS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; METALS; MOCKUP; NATIONAL ORGANIZATIONS; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIATION DETECTORS; RADIOISOTOPES; RESPIRATORY SYSTEM; SEMICONDUCTOR DETECTORS; SIMULATION; SORPTION; SPONTANEOUS FISSION RADIOISOTOPES; STRUCTURAL MODELS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 10 refs.