Quantitative WDS analysis using electron probe microanalyzer

  • 1. Research Institute, King Fahd University of Petroleum and Minerals, P.O. Box 1073, Dhahran 31261 (Saudi Arabia)
  • 2. Saudi Aramco, P.O. Box 65, Tanajib 31311 (Saudi Arabia)

Description

In this paper, the procedure for conducting quantitative elemental analysis by ZAF correction method using wavelength dispersive X-ray spectroscopy (WDS) in an electron probe microanalyzer (EPMA) is elaborated. Analysis of a thermal barrier coating (TBC) system formed on a Ni-based single crystal superalloy is presented as an example to illustrate the analysis of samples consisting of a large number of major and minor elements. The analysis was performed by known standards and measured peak-to-background intensity ratios. The procedure for using separate set of acquisition conditions for major and minor element analysis is explained and its importance is stressed

Additional details

Identifiers

DOI
10.1016/j.matchar.2005.11.007;
PII
S1044-5803(05)00256-1;

Publishing Information

Journal Title
Materials Characterization
Journal Volume
56
Journal Issue
3
Journal Page Range
p. 192-199
ISSN
1044-5803
CODEN
MACHEX

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38007608
Subject category
S36: MATERIALS SCIENCE;
Descriptors DEI
ELECTRON MICROPROBE ANALYSIS; ELECTRON PROBES; MONOCRYSTALS; WAVELENGTHS; X-RAY SPECTROSCOPY
Descriptors DEC
CHEMICAL ANALYSIS; CRYSTALS; MICROANALYSIS; NONDESTRUCTIVE ANALYSIS; PROBES; SPECTROSCOPY

Optional Information

Copyright
Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.