Published February 1983 | Version v1
Journal article

A fundamental approach to the design of a dose-rate calculation program for use in brachytherapy planning

Creators

  • 1. Saint Luke's Hospital, Guildford (UK)

Description

A method, developed from the Quantisation Method, of calculating dose-rate distributions around uniformly and non-uniformly loaded brachytherapy sources is described. It allows accurate and straightforward corrections for oblique filtration and self-absorption to be made. Using this method, dose-rate distributions have been calculated for sources of radium 226, gold 198, iridium 192, caesium 137 and cobalt 60, all of which show very good agreement with existing measured and calculated data. This method is now the basis of the Interstitial and Intracavitary Dosimetry (IID) program on the General Electric RT/PLAN computerised treatment planning system. (author)

Additional details

Publishing Information

Journal Title
Br. J. Radiol.
Journal Volume
56
Journal Issue
662
Series
Br. J. Radiol.
Journal Page Range
113-119