Mapping of isoexposure curves for evaluation of equivalent environmental doses for radiodiagnostic mobile equipment
Creators
- 1. Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre, RS (Brazil). Hospital de Clinicas. Setor de Fisica Medica e Radioprotecao
- 2. Irmandade da Santa Casa de Misericordia de Porto Alegre, RS (Brazil). Servico de Atencao a Saude e Qualidade de Vida
- 3. Pontificia Univ. Catolica do Rio Grande do Sul (PUC/RS), Porto Alegre, RS (Brazil). Grupo de Experimentacao e Simulacao Computacional em Fisica Medica
Description
This paper generates iso exposure curves in areas where the mobile radiodiagnostic equipment are used for evaluation of iso kerma map and the environment equivalent dose (H*(d)). It was used a Shimadzu mobile equipment and two Siemens, with non anthropomorphic scatter. The exposure was measured in a mesh of 4.20 x 4.20 square meter in steps of 30 cm, at half height from the scatterer. The calculation of H*(d) were estimated for a worker present in all the procedures in a period of 11 months, being considered 3.55 m As/examination and 44.5 procedures/month (adult UTI) and 3.16 m As/examination and 20.1 procedure/month (pediatric UTI), and 3.16 m As/examination and 20.1 procedure/month (pediatric UTI). It was observed that there exist points where the H*(d) was over the limit established for the free area inside the radius of 30 cm from the central beam of radiation in the case of pediatric UTI and 60 cm for adult UTI. The points localized 2.1 m from the center presented values lower than 25% of those limit
Files
42081725.pdf
Files
(773.8 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:d5729bff4241d7f293a963ea3c0b7974
|
773.8 kB | Preview Download |
Additional details
Additional titles
- Original title (Portuguese)
- Mapeamento de curvas de isoexposicao para avaliacao de equivalente de dose ambiente para equipamentos moveis de radiodiagnostico
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- INIS-BR--9608
Conference
- Title
- 2. Congress on radioprotection of the Portuguese language countries; 2. Brazilian congress on radiation protection; 5. International congress on industrial radioprotection
- Original Conference Title
- Radio 2011: 2. Congresso de protecao contra radiacoes da Comunidade dos Paises de Lingua Portuguesa
- Acronym
- Radio 2011
- Dates
- 28-31 May 2010
- Place
- Recife, PE (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 42081725
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DIAGNOSIS; DOSE COMMITMENTS; DOSE EQUIVALENTS; DOSIMETRY; GAMMA RADIATION; MEDICINE; NUCLEAR MEDICINE; PERSONNEL MONITORING; RADIATION DOSES; RADIATIONS; RADIOLOGY; X RADIATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; MONITORING; NUCLEAR MEDICINE; RADIATION MONITORING; RADIATIONS; RADIOLOGY