Reproducibilities of the successive first pass radionuclide angiography using pertechnatate-99m

  • 1. Tokyo Metropolitan Geriatric Medical Center (Japan)

Description

Reproducibilities of the successive first pass radionuclide angiography using pertechnatate-99m was evaluated. Ten minutes after the initial study of the first pass angiography, successive study was performed at the same resting condition. The LVEF showed good correlation between initial study and successive study. By using slant hole collimator attached close to the chest anteriory, left ventricle was separated from the descending artery completely. Multigated image was obtained by 8 beats summation in this condition. Good correlation was obtained between 8 beats method and 4 beats method and the eight beats method produced image with better configuration and improvement in statistics. (author)

Additional details

Publishing Information

Journal Title
Kaku Igaku
Journal Volume
18
Journal Issue
10
Series
Kaku Igaku.
Journal Page Range
1399-1404
ISSN
0022-7854