We describe a method for calibration and test of large-area TOF counters using cosmic radiation. We applied the method to the time-of-flight system of the DELPHI detector at the LEP e+e- storage ring, made of scintillation (NE110) counters (20x350 cm2). The photomultipliers used (EMI 9902KB) reach an average gain of 5x108 at 1700 V and the time resolution achieved is 1.2 ns. Using this method we measured the counter efficiencies as a function of the position; we obtained 135 cm for the effective attenuation length and 40 photoelectrons for a minimum-ionizing particle crossing the center of the counter. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research, Section A
- Journal Volume
- 292
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 319-328
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 21083948
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ATTENUATION; CALIBRATION; EFFICIENCY; PERFORMANCE TESTING; PLASTIC SCINTILLATION DETECTOR; TIME RESOLUTION; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTROMETERS; TESTING; TIMING PROPERTIES