Calibration for radiation protection survey meter SSM1
Description
The calibration functions for the Radiation Protection Survey Meter SSM1 built-in Geiger Mueller measuring tubes of the type ZP 1201 (low dose rate tube) and ZP 1313 (high dose rate tube) are given in the present study. Six dose rate points are proposed as the calibration points of covering the range from 500 nSv/h up to 5 Sv/h. The difference between the true dose rate and the values calculated with the method in this paper for No. 382 SSM1 is less than ±2%. (author)
Availability note (English)
Available from the Austrian Research Centre Seibersdorf, A-2444 Seibersdorf (AT).Additional details