Published March 1992 | Version v1
Report

Calibration for radiation protection survey meter SSM1

Description

The calibration functions for the Radiation Protection Survey Meter SSM1 built-in Geiger Mueller measuring tubes of the type ZP 1201 (low dose rate tube) and ZP 1313 (high dose rate tube) are given in the present study. Six dose rate points are proposed as the calibration points of covering the range from 500 nSv/h up to 5 Sv/h. The difference between the true dose rate and the values calculated with the method in this paper for No. 382 SSM1 is less than ±2%. (author)

Availability note (English)

Available from the Austrian Research Centre Seibersdorf, A-2444 Seibersdorf (AT).

Additional details

Publishing Information

Imprint Pagination
30 p.
Report number
OEFZS-A--2225

INIS

Country of Publication
Austria
Country of Input or Organization
Austria
INIS RN
24016054
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Non-conventional Literature
Descriptors DEI
CALIBRATION; RADIATION MONITORS
Descriptors DEC
MEASURING INSTRUMENTS; MONITORS

Optional Information

Notes
Limited distribution.