Published September 2001
| Version v1
Journal article
The response of high-purity germanium detectors to X-rays with energy in the region of the Ge K-absorption edge
Creators
Description
The response of a high-purity germanium detector to X-rays in the 8-15-keV energy region has been investigated. The w-value and energy resolution dependencies on the X-ray energy have been studied. No abrupt variation of w is observed at the germanium K absorption-edge (11.104-keV). The detector energy resolution follows a characteristic linear dependence on Ex-1/2 over the whole energy range. Both features are consistent with the absence of intrinsic non-linearity effects in germanium at the K-edge
Additional details
Identifiers
- PII
- S0969804301000586;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 55
- Journal Issue
- 3
- Journal Page Range
- p. 331-334
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Canada
- INIS RN
- 33030291
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ENERGY RANGE; ENERGY RESOLUTION; HIGH-PURITY GE DETECTORS; X-RAY SOURCES
- Descriptors DEC
- GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATION SOURCES; RESOLUTION; SEMICONDUCTOR DETECTORS
Optional Information
- Copyright
- Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.