Published September 2001 | Version v1
Journal article

The response of high-purity germanium detectors to X-rays with energy in the region of the Ge K-absorption edge

Description

The response of a high-purity germanium detector to X-rays in the 8-15-keV energy region has been investigated. The w-value and energy resolution dependencies on the X-ray energy have been studied. No abrupt variation of w is observed at the germanium K absorption-edge (11.104-keV). The detector energy resolution follows a characteristic linear dependence on Ex-1/2 over the whole energy range. Both features are consistent with the absence of intrinsic non-linearity effects in germanium at the K-edge

Additional details

Identifiers

PII
S0969804301000586;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
55
Journal Issue
3
Journal Page Range
p. 331-334
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.