Description

The design, construction and test of a charged particle detector made of scintillation counters read by Silicon Photomultipliers (SiPM) is described. The detector, which operates in vacuum and is used as a veto counter in the NA62 experiment at CERN, has a single channel time resolution of 1.14 ns, a spatial resolution of ∼2.5 mm and an efficiency very close to 1 for penetrating charged particles

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/11/03/P03029

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
11
Journal Issue
03
Journal Page Range
p. P03029
ISSN
1748-0221

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47072207
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CERN; CHARGED PARTICLES; DESIGN; EFFICIENCY; PHOTOMULTIPLIERS; SCINTILLATION COUNTERS; SILICON; SPATIAL RESOLUTION; TIME RESOLUTION
Descriptors DEC
ELEMENTS; INTERNATIONAL ORGANIZATIONS; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION; SEMIMETALS; TIMING PROPERTIES