Published March 2016
| Version v1
Journal article
CHANTI: a fast and efficient charged particle veto detector for the NA62 experiment at CERN
Creators
- 1. Dipartimento di Fisica ''Ettore Pancini'', Università degli Studi di Napoli Federico II, via Cinthia, Napoli (Italy)
- 2. INFN Sezione di Napoli, via Cinthia, Napoli (Italy)
- 3. INFN Laboratori Nazioni di Frascati, Via E. Fermi 40, Frascati (Italy)
Description
The design, construction and test of a charged particle detector made of scintillation counters read by Silicon Photomultipliers (SiPM) is described. The detector, which operates in vacuum and is used as a veto counter in the NA62 experiment at CERN, has a single channel time resolution of 1.14 ns, a spatial resolution of ∼2.5 mm and an efficiency very close to 1 for penetrating charged particles
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/11/03/P03029Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 11
- Journal Issue
- 03
- Journal Page Range
- p. P03029
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47072207
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CERN; CHARGED PARTICLES; DESIGN; EFFICIENCY; PHOTOMULTIPLIERS; SCINTILLATION COUNTERS; SILICON; SPATIAL RESOLUTION; TIME RESOLUTION
- Descriptors DEC
- ELEMENTS; INTERNATIONAL ORGANIZATIONS; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION; SEMIMETALS; TIMING PROPERTIES