Soft x-ray pulse height analyzer in the HT-7 tokamak
Creators
- 1. Institute of Plasma Physics, Chinese Academy of Sciences, P.O. Box 1126, Hefei 230031 (China)
Description
A new soft x-ray pulse height analyzer (PHA), based on a Silicon Drift Detector (SDD) linear array consisting of 15 SDD detectors, has been installed in the HT-7 tokamak. The energy resolution of 150-180 eV at photon energy of 6 keV is achieved for the SDD with Peltier cooling. The effective time response of the SDD PHA is 50 ms. The profiles of electron temperature and the intensity of metallic impurities can be obtained with a spatial resolution of 3 cm. The performance and first experimental results from the new PHA system are presented
Additional details
Identifiers
- DOI
- 10.1063/1.1808911;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 75
- Journal Issue
- 11
- Journal Page Range
- p. 4930-4933
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36085030
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- COOLING; ELECTRON TEMPERATURE; ENERGY RESOLUTION; EV RANGE 100-1000; HT-7 TOKAMAK; ION TEMPERATURE; KEV RANGE 01-10; PHOTONS; PLASMA DIAGNOSTICS; PLASMA IMPURITIES; PULSES; SI SEMICONDUCTOR DETECTORS; SILICON; SOFT X RADIATION; SPATIAL RESOLUTION; THERMONUCLEAR REACTORS
- Descriptors DEC
- BOSONS; CLOSED PLASMA DEVICES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ELEMENTS; ENERGY RANGE; EV RANGE; IMPURITIES; IONIZING RADIATIONS; KEV RANGE; MASSLESS PARTICLES; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; RESOLUTION; SEMICONDUCTOR DETECTORS; SEMIMETALS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES; X RADIATION
Optional Information
- Notes
- (c) 2004 American Institute of Physics