Published March 2006 | Version v1
Journal article

Design, construct and test a calibration radon chamber

  • 1. Atomic Energy Commission, Damascus (Syrian Arab Republic), Dept. of Protection and Safety

Description

a new radon chamber has been designed, constructed and tested. The design was chosen to meet the needs of several different radon detection techniques, both active and passive. The chamber is a cubic shape made of pixy glass with volume about 0.65 m3. On top, it has a circular opening with double lock system for entering and removing samples without significant disturbance of the inside concentration. It also has the possibility to expose passive radon detectors at different levels inside the chamber. In addition, two fans, with variable speeds, were fitted in two opposite corners to maintain the uniformity of radon gas inside the volume. Many other features were also fitted to it. This calibration chamber can be used to perform researches and studies on radon behaviors, in addition to calibrate different types of detectors (passive and active) which are used for measuring radon and its daughters. Solid radium-226 source with activity of about 12 kBq was placed at the bottom of the chamber to generate radon. (author)

Additional details

Publishing Information

Journal Title
Aalam Al-Zarra
Journal Issue
102
Journal Page Range
p. 79
ISSN
1607-985X

Optional Information

Notes
Abstract of Scientific Research