Published May 2020 | Version v1
Journal article

Coincidence summing corrections factors calculated for volume 152Eu sources in γ-ray spectrometry using Monte Carlo techniques

  • 1. Faculte de Physique, Laboratoire SNIRM, USTHB, Algiers (Algeria)
  • 2. Centre de Recherche Nucleaire d'Alger, Algiers (Algeria)

Description

True coincidence summing (TCS) correction factors were determined for 152Eu volume sources used for γ-ray efficiency calibration as the ratio of full energy peak efficiency (ε) simulated by MCNP5 code to the apparent full energy peak efficiency (εapp) determined experimentally. The results were compared to those calculated by TrueCoinc software and then validated using certified environmental radioactive samples. The dependence of 152Eu and 133Ba TCS factors on volume sample and matrix was investigated. Finally, as our n-type HPGe detector is very sensitive to low energy photons, the contribution of the γ-X coincidence summing to TCS factors for 152Eu and 133Ba sources in volume soil samples was studied. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
324
Journal Issue
2
Journal Page Range
p. 687-696
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
35 refs.