Published May 12, 2010 | Version v1
Report

Development of a Time-resolved Soft X-ray Spectrometer for Laser Produced Plasma Experiments

Description

A 2400 line/mm variable spaced grating spectrometer (VSG) has been used to measure soft x-ray emission (8-22 (angstrom)) from laser-produced plasma experiments at Lawrence Livermore National Laboratory's Compact Multipulse Terrawatt (COMET) Laser Facility. The spectrometer was coupled to a Kentech x-ray streak camera to study the temporal evolution of soft x-rays emitted from the back of mylar and copper foils irradiated at 1015 W/cm2. The instrument demonstrated a resolving power of ∼ 120 at 19 (angstrom) with a time resolution of 31 ps. The time-resolved copper emission spectrum was consistent with a photodiode monitoring the laser temporal pulse shape and indicated that the soft x-ray emission follows the laser heating of the target. The time and spectral resolution of this diagnostic make it useful for studies of high temperature plasmas.

Availability note (English)

Available from https://e-reports-ext.llnl.gov/pdf/396790.pdf; PURL: https://www.osti.gov/servlets/purl/1009648-fu3qjg/

Additional details

Publishing Information

Imprint Pagination
5 p.
Report number
LLNL-CONF--431867

Conference

Title
High Temperature Plasma Diagnostics 2010
Dates
16-20 May 2010
Place
Wildwood, NJ (United States)

Optional Information

Contract/Grant/Project number
W-7405-ENG-48
Notes
PDF-FILE: 5; SIZE: 0.5 MBYTES
Funding organization
US Department of Energy (United States)