Published 2005 | Version v1
Journal article

Assessment of mammography spectra using compton spectrometry techniques

  • 1. Departamento de Ingenieria Quimica y Nuclear, Universidad Politecnica de Valencia, Apartado 22012, E-46071 Valencia (Spain)
  • 2. Servicio de Proteccion Radiologica, Hospital La Fe de Valencia, Avenida Campanar, 46016 Valencia (Spain)

Description

The quality control of mammography units is necessary to reduce the dose imparted to women as much as possible. An accurate characterisation of the primary X-ray spectra is very useful for this purpose. Primary spectra can be obtained using Compton spectrometry techniques. In this work, a commercial spectrometer used to characterise a mammography X-ray tube has been simulated using the Monte Carlo method by means of the MCNP code. Using the developed model, a Response matrix is obtained. Owing to the fact that this matrix is ill-conditioned, the inversion is not a simple process. This problem has been solved using the truncated singular value decomposition method. Results obtained when this methodology was applied have been compared with theoretical X-ray spectra. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci243

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
116
Journal Issue
1-4
Journal Page Range
p. 627-630
ISSN
0144-8420

Conference

Title
10. International Conference on Radiation Shielding, and 13. ANS Topical Meeting on Radiation Protection and Shielding - ICRS-10/RPS 2004
Dates
9-14 May 2004
Place
Funchal, Madeira Island (Portugal)

Optional Information

Notes
13 refs