Published October 1979 | Version v1
Journal article

Determination of time constants of reactor pressure and temperature sensors: the dynamic data system method

  • 1. Univ. of Wisconsin, Madison

Description

A new modeling technique is introduced for on-line sensor time constant identification, both for the resistance temperature detector (RTD) and for the pressure sensor using power plant operational data. The sensor's time constant is estimated from a real characteristic root of the fitted autoregressive moving average model. The RTD's time constant values were identified to be 8.4 s, with a standard deviation of 1.2 s. The pressure sensor time constant was identified to be 28.6 ms, with a standard deviation of 3.5 ms

Additional details

Publishing Information

Journal Title
Nucl. Sci. Eng.
Journal Volume
72
Journal Issue
1
Series
Nucl. Sci. Eng.
Journal Page Range
84-96
ISSN
0029-5639

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
11538454
Subject category
S22: GENERAL STUDIES OF NUCLEAR REACTORS;
Descriptors DEI
MATHEMATICAL MODELS; ON-LINE SYSTEMS; PRESSURE GAGES; REACTOR COMPONENTS; SENSITIVITY; TEMPERATURE MONITORING; TIME DEPENDENCE
Descriptors DEC
MEASURING INSTRUMENTS; MONITORING