Published February 2001
| Version v1
Report
Evaluation of the analytical errors in plutonium and uranium isotopic composition analysis based on routine control measurement data (1998/9 - 2000/12)
Creators
- 1. Japan Nuclear Cycle Development Inst., Tokai, Ibaraki (Japan)
- 2. Nuclear Technology Engineering Co. Ltd., Tokai, Ibaraki (Japan)
- 3. Inspection Development Co. Ltd., Tokai, Ibaraki (Japan)
Description
Four mass spectrometers are used for plutonium and uranium isotopic composition analysis in the Plutonium Fuel Center. Analytical errors of the analysis was evaluated based on routine control data obtained with measurement of reference materials. It was confirmed that the errors satisfied the International Target Values for safeguard analysis, and the random error of the analysis was improved by changing conventional method to total evaporation method. (author)
Availability note (English)
Available from JICST Library (JICST: Japan Science and Technology Corporation, Information Center for Science and Technology), P.O. Box 10 Hikarigaoka, Tokyo 179-9810 Japan, FAX: +81-3-3979-4781, JICST Service Homepage: www.jst.go.jp/EN/Additional details
Publishing Information
- Imprint Pagination
- 36 p.
- Report number
- JNC-TN--8440-2001-005
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 33003638
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Non-conventional Literature, Numerical Data
- Descriptors DEI
- CONTROL; ERRORS; EVALUATED DATA; EVAPORATION; EXPERIMENTAL DATA; ISOTOPE RATIO; MASS SPECTROMETERS; MASS SPECTROSCOPY; PLUTONIUM; URANIUM
- Descriptors DEC
- ACTINIDES; DATA; ELEMENTS; INFORMATION; MEASURING INSTRUMENTS; METALS; NUMERICAL DATA; PHASE TRANSFORMATIONS; SPECTROMETERS; SPECTROSCOPY; TRANSURANIUM ELEMENTS
Optional Information
- Notes
- 16 figs., 9 tabs., 1 photo.