Published February 2011 | Version v1
Journal article

Computer-Aided Evaluation of Breast MRI for the Residual Tumor Extent and Response Monitoring in Breast Cancer Patients Receiving Neoadjuvant Chemotherapy

  • 1. Seoul National University Hospital and the Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul (Korea, Republic of)
  • 2. Seoul National University Bundang Hospital, Seongnam (Korea, Republic of)
  • 3. Hanyang University College of Medicine, Hanyang University Hospital, Seoul (Korea, Republic of)
  • 4. School of Medicine, Ewha Womans University, Seoul (Korea, Republic of)

Description

To evaluate the accuracy of a computer-aided evaluation program (CAE) of breast MRI for the assessment of residual tumor extent and response monitoring in breast cancer patients receiving neoadjuvant chemotherapy. Fifty-seven patients with breast cancers who underwent neoadjuvant chemotherapy before surgery and dynamic contrast enhanced MRI before and after chemotherapy were included as part of this study. For the assessment of residual tumor extent after completion of chemotherapy, the mean tumor diameters measured by radiologists and CAE were compared to those on histopathology using a paired student t-test. Moreover, the agreement between unidimensional (1D) measurement by radiologist and histopathological size or 1D measurement by CAE and histopathological size was assessed using the Bland-Altman method. For chemotherapy monitoring, we evaluated tumor response through the change in the 1D diameter by a radiologist and CAE and three-dimensional (3D) volumetric change by CAE based on Response Evaluation Criteria in Solid Tumors (RECIST). Agreement between the 1D response by the radiologist versus the 1D response by CAE as well as by the 3D response by CAE were evaluated using weighted kappa (k) statistics. For the assessment of residual tumor extent after chemotherapy, the mean tumor diameter measured by radiologists (2.0 ± 1.7 cm) was significantly smaller than the mean histological diameter (2.6 ± 2.3 cm) (p = 0.01), whereas, no significant difference was found between the CAE measurements (mean = 2.2 ± 2.0 cm) and histological diameter (p = 0.19). The mean difference between the 1D measurement by the radiologist and histopathology was 0.6 cm (95% confidence interval: -3.0, 4.3), whereas the difference between CAE and histopathology was 0.4 cm (95% confidence interval: -3.9, 4.7). For the monitoring of response to chemotherapy, the 1D measurement by the radiologist and CAE showed a fair agreement (k = 0.358), while the 1D measurement by the radiologist and 3D measurement by CAE showed poor agreement (k = 0.106). CAE for breast MRI is sufficiently accurate for the assessment of residual tumor extent in breast cancer patients receiving neoadjuvant chemotherapy. However, for the assessment of response to chemotherapy, the assessment by the radiologist and CAE showed a fair to poor agreement

Additional details

Publishing Information

Journal Title
Korean Journal of Radiology
Journal Volume
12
Journal Issue
1
Series
31 refs, 3 figs, 3 tabs
Journal Page Range
p. 34-43
ISSN
1229-6929

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
42044926
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CHEMOTHERAPY; CONTRAST MEDIA; MAMMARY GLANDS; MONITORING; NEOPLASMS; NMR IMAGING; PATIENTS; SURGERY
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MEDICINE; ORGANS; THERAPY