Published January 15, 1989 | Version v1
Journal article

NMR measurements on the TRIUMF polarized deuterium target

  • 1. British Columbia Univ., Vancouver (Canada). TRIUMF Facility

Description

The NMR system and the polarization measurements of the TRIUMF polarized deuterium target are described. The NMR tuned circuit is kept resonant at all frequencies to allow for a high gain in the rf amplifiers and to permit the measurement of the real part of the NMR signal. A Starburst (J-11) and LSI-11 are used to process the NMR signal at a high sample rate. A synthesizer under the control of an 8085 based microprocessor (TRIMAC) provides the rf sweep. Vector polarizations between -0.33 and -0.40 were routinely achieved during a four week run in the TRIUMF M11 pion beamline with a statistical uncertainty of 3.4% and a systematic uncertainty of 5%. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
274
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
515-521
ISSN
0168-9002
CODEN
NIMAE