Published June 2008 | Version v1
Journal article

Calibration of a radionuclide calibrator system as a Bulgarian standard for activity

  • 1. Physikalisch-Technische Bundesanstalt, Department 6.1, Bundesallee 100, D-38116 Braunschweig (Germany)
  • 2. National Centre of Metrology, G.M. Dimitrov Blvd. 52B, 1040 Sofia (Bulgaria)

Description

A commercially available radionuclide calibrator of the type Curiementor 3 (PTW, Freiburg, Germany) was calibrated using activity standard solutions, from the Physikalisch-Technische Bundesanstalt (PTB), of a number of photon-emitting radionuclides, in particular those used in nuclear medicine: 18F, 51Cr, 67Ga, 99mTc, 111In, 123I, 125I, 131I, 186Re, 201Tl, and for beta-emitting nuclides: 32P, 89Sr, 90Y, 204Tl. An energy-dependent efficiency curve was established to calculate efficiencies for photon-emitting radionuclides. The differences between experimental and calculated radionuclide efficiencies of the measuring system are of the order of a few percent. The performance of the system was checked for instrument stability, using a long-lived 137Cs source, and for linearity using the decay of a 99mTc source. The calibration measurements were carried out with 2 g of solution in standard PTB-type glass ampoules. The dependence of efficiency with varying solution mass and changing ampoule geometry is studied for some of the radionuclides. For example, geometry correction factors are determined for various solution volumes or ampoules with different wall thickness. The measuring system will enable the National Centre of Metrology (NCM) of Bulgaria to provide calibration services for activity measurements, which are traceable to PTB standards, and to organize national comparisons in order to improve the quality of measurements for nuclear medicine in Bulgarian hospitals. Furthermore, it should enable NCM to declare Calibration and Measurement Capabilities (CMCs) and to submit radioactive solutions to the International Reference System (SIR) at the BIPM in order to compare the results with Key Comparison Reference Values (KCRVs)

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2008.02.042

Additional details

Identifiers

DOI
10.1016/j.apradiso.2008.02.042;
PII
S0969-8043(08)00093-6;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
66
Journal Issue
6-7
Journal Page Range
p. 965-971
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
16. international conference on radionuclide metrology and its applications
Dates
3-7 Sep 2007
Place
Cape Town (South Africa)

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.