Published April 2000 | Version v1
Journal article

A comparison study of linear reconstruction techniques for diffuse optical tomographic imaging of absorption coefficient

  • 1. CenSISS, CDSP Center, 442 DA, Department of Electrical and Computer Engineering, Northeastern University, Boston, MA 02115 (United States)
  • 2. CenSISS, CER, 235 Forsyth Building, Department of Electrical and Computer Engineering, Northeastern University, Boston, MA 02115 (United States)
  • 3. Department of Mathematics, Tufts University, Boston, MA (United States)
  • 4. CIMIT and Mass General Hospital NMR Center, Harvard Medical School, Boston, MA (United States)

Description

We compare, through simulations, the performance of four linear algorithms for diffuse optical tomographic reconstruction of the three-dimensional distribution of absorption coefficient within a highly scattering medium using the diffuse photon density wave approximation. The simulation geometry consisted of a coplanar array of sources and detectors at the boundary of a half-space medium. The forward solution matrix is both underdetermined, because we estimate many more absorption coefficient voxels than we have measurements, and ill-conditioned, due to the ill-posedness of the inverse problem. We compare two algebraic techniques, ART and SIRT, and two subspace techniques, the truncated SVD and CG algorithms. We compare three-dimensional reconstructions with two-dimensional reconstructions which assume all inhomogeneities are confined to a known horizontal slab, and we consider two 'object-based' error metrics in addition to mean square reconstruction error. We include a comparison using simulated data generated using a different FDFD method with the same inversion algorithms to indicate how our conclusions are affected in a somewhat more realistic scenario. Our results show that the subspace techniques are superior to the algebraic techniques in localization of inhomogeneities and estimation of their amplitude, that two-dimensional reconstructions are sensitive to underestimation of the object depth, and that an error measure based on a location parameter can be a useful complement to mean squared error. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
45
Journal Issue
4
Journal Page Range
p. 1051-1070
ISSN
1361-6560

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
31022278
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; COMPUTERIZED TOMOGRAPHY; DEPTH; IONOGRAPHIC IMAGING; LINEAR PROGRAMMING; THREE-DIMENSIONAL CALCULATIONS; TWO-DIMENSIONAL CALCULATIONS
Descriptors DEC
BIOMEDICAL RADIOGRAPHY; DIAGNOSTIC TECHNIQUES; DIMENSIONS; MEDICINE; PROGRAMMING; TOMOGRAPHY