Wide-range structurally optimized channel for monitoring the certified power of small-core reactors
- 1. Russian Federal Nuclear Center—All-Russian Scientific Research Institute of Experimental Physics (Russian Federation)
Description
The results of tests of a prototype version of a channel for monitoring the certified power of small-core reactors performed at the BR-K1 reactor at the All-Russian Scientific Research Institute of Experimental Physics are reported. An SNM-11 counter and commercial KNK-4 and KNK-3 compensated ion chambers were used as neutron detectors in the tested channel, and certified NCMM and CCMM measurement modules controlled by a PC with specialized software were used as measuring instruments. The specifics of metrological assurance of calibration of the channel in the framework of reactor power monitoring are discussed.
Additional details
Identifiers
Publishing Information
- Journal Title
- Physics of Atomic Nuclei
- Journal Volume
- 79
- Journal Issue
- 8
- Journal Page Range
- p. 1349-1356
- ISSN
- 1063-7788
- CODEN
- PANUEO
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48059063
- Subject category
- S22: GENERAL STUDIES OF NUCLEAR REACTORS;
- Descriptors DEI
- CALIBRATION; COMPUTER CODES; IONIZATION CHAMBERS; MONITORING; NEUTRON DETECTORS; POWER DENSITY; REACTOR CORES
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS; REACTOR COMPONENTS
Optional Information
- Copyright
- Copyright (c) 2016 Pleiades Publishing, Ltd.