Establishment of a primary standard system for low energy X-rays using a free air ionization chamber

Description

In this work a primary standard system was established for low energy X-rays (10 kV to 50 kV), using a free air ionization chamber with concentric cylinders, Victoreen (Model 481-5), at the Calibration Laboratory of Instruments (LCI) of the Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP). For this, a new ionization chamber alignment protocol was developed for the radiation system and a modification on the micrometer housing used for the movement of the internal cylinders was ma de. The results obtained for the stability and characterization tests showed to be within the limits established by the standards IEC 61674 and IEC 60731. The correction factors for photon attenuation in the air, transmission and scattering in the diaphragm, scattering and fluorescence and ion recombination were also determined. These values were compared with those obtained by the German primary standard laboratory, Physikalisch-Technische Bundesanstalt (PTB), showing good agreement. Finally, the absolute values of the quantity air kerma rate for the standard qualities direct beams MWV28 and WMV35 and the attenuated beams WMH28 and WMH35 were determined; the results are in agreement, with a maximum difference of 3,8% with the values obtained using the secondary standard system of LCI. (author)

Files

47088800.pdf

Files (6.9 MB)

Name Size Download all
md5:ea36766141ddcd78d16b0a6fb080b7eb
6.9 MB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Estabelecimento de um sistema padrao primario para raios-X de energias baixas com uma camara de ionizacao de ar livre

Publishing Information

Imprint Pagination
114 p.
Report number
INIS-BR--16508