Simulator films and port films are important documents of quality assurance of treatment delivery. Graticules have been constructed for linear accelerators which project mutually perpendicular arrays of equi-spaced dots on X ray films, similar to those projected by a graticule on the simulator. The spacing corresponds to 2 cm at the isocentric distance. The projected dot arrays function as built-in rulers in the films and greatly facilitate quantitative comparison of simulator films and treatment verification films, regardless of film magnification