Published August 1987 | Version v1
Journal article

The simplified method of measuring radon in mineral waters by the liquid scintillation counter

  • 1. Okayama Univ., Misasa, Tottori (Japan). Misasa Hospital

Description

Radon in mineral waters was determined with simplicity by newly developed method. In this method, 2.0 ml of water sample was directly mixed with 15 ml of Hionic-Fluor in a vial, which was then filled up with additional Hionic-Fluor; 190 minutes were allowed for the radioactivity to equilibrate. The radioactivity of Radon was measured by the liquid Scintillation Counter (Tri-Carb 460C). Radon content was easily calculated with the proposed formula by an electric desk calculator. The reproducibility was proved CV = 5.77 %; the lowest detection limit was 0.2 nCi-1. An adequate correlation was obtained from the results of measurements compared with the original method; r = 0.996, y = 0.818x x 0.875. (author)

Additional details

Publishing Information

Journal Title
Kankyo Byotaiken Hokoku
Journal Issue
no.58
Series
Kankyo Byotaiken Hokoku.
ISSN
0913-3771
CODEN
KBHOE