Verification of a pencil beam based treatment planning system: output factors for open photon beams shaped with MLC or blocks
Creators
- 1. Department of Medical Physics and Bioengineering, Maelarsjukhuset, SE-631 88 Eskilstuna (Sweden).
- 2. Department of Radiation Physics, Lund University Hospital, SE-221 85 Lund (Sweden)
Description
The accuracy of monitor unit calculations from a pencil beam based, three-dimensional treatment planning system (3D TPS) has been evaluated for open irregularly shaped photon fields. The dose per monitor unit was measured in water and in air for x-ray beam qualities from 6 to 15 MV. The fields were shaped either with a multileaf collimator (MLC) or with customized alloy blocks. Calculations from the 3D TPS were compared with measurements. The agreement between calculated and measured dose per monitor unit depended on field size and the amount of blocking and was within 3% for the MLC-shaped fields. The deviation could be traced to limitations in head scatter modelling for the MLC. For fields shaped with alloy blocks, the dose per monitor unit was calculated to be within 1.6% of measured values for all fields studied. The measured and calculated relative phantom scatter for fields with the same equivalent field size were identical for MLC and alloy shaped fields. These results indicate that the accuracy in the TPS calculations for open irregular fields, shaped with MLC or blocks, is satisfactory for clinical situations. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 44
- Journal Issue
- 9
- Journal Page Range
- p. N201-N207
- ISSN
- 1361-6560
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 31025417
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BEAM INJECTION; BEAM MONITORING; CALCULATION METHODS; RADIATION DOSES; RADIOTHERAPY; SCATTERING
- Descriptors DEC
- MEDICINE; MONITORING; THERAPY
Optional Information
- Notes
- Refs