Quality assurance procedure for assessing mechanical accuracy of a radiation field center in stereotactic radiotherapy
Creators
- 1. Osaka City Univ., Hospital, Osaka, Osaka (Japan)
- 2. Osaka City Univ., Graduate School of Medicine, Osaka, Osaka (Japan)
Description
Stereotactic radiotherapy requires a quality assurance (QA) program that ensures the mechanical accuracy of a radiation field center. We have proposed a QA method for achieving the above requirement by conducting the Winston Lutz test using an electronic portal image device (EPID). An action limit was defined as three times the standard deviation. Then, the action limits for mean deviations of the radiation field center during collimator rotation, gantry rotation, and couch rotation in clockwise and counterclockwise resulted in 0.11 mm, 0.52 mm, 0.37 mm, and 0.41 mm respectively. Two years after the QA program was launched, the mean deviation of the radiation field center during gantry rotation exceeded the above action limit. Consequently, a mechanical adjustment for the gantry was performed, thereby restoring the accuracy of the radiation field center. A field center shift of 0.5 mm was also observed after a micro multi-leaf collimator was unmounted. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 68
- Journal Issue
- 10
- Journal Page Range
- p. 1333-1339
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 44034064
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BALL BEARINGS; COLLIMATORS; IMAGE PROCESSING; LINEAR ACCELERATORS; QUALITY ASSURANCE; RADIOTHERAPY; ROTATION; X RADIATION
- Descriptors DEC
- ACCELERATORS; BEARINGS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; MOTION; NUCLEAR MEDICINE; PROCESSING; RADIATIONS; RADIOLOGY; THERAPY