Published December 1975
| Version v1
Journal article
Calorimetry automation in the 0.4 to 3 K range: use of the NBS superconductive fixed point device and the specific heat of copper
Description
Operation of an automatic data acquisition system, with an on-line minicomputer, in the 0.4 to 30K temperature range is described and illustrated by results obtained on pure copper. The temperature scale was set up using both 3He vapor pressure thermometry and the NBS superconductive fixed point device. On the basis of specific heat measurements on pure Cu and Cu alloys, it appears that the temperatures assigned by the NBS to the superconducting transition temperatures of cadmium and zinc are too close. The required adjustment is within their assigned uncertainty limits
Additional details
Identifiers
- DOI
- 10.1063/1.1134138;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 46
- Journal Issue
- 12
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1670-1675
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 7239831
- Subject category
- S36: MATERIALS SCIENCE; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- CALORIMETRY; COPPER; COPPER ALLOYS; DATA ACQUISITION; SPECIFIC HEAT; TEMPERATURE MEASUREMENT; ULTRALOW TEMPERATURE
- Descriptors DEC
- ALLOYS; ELEMENTS; METALS; PHYSICAL PROPERTIES; THERMODYNAMIC PROPERTIES; TRANSITION ELEMENT ALLOYS; TRANSITION ELEMENTS
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent