Description

Intensity-modulated radiation therapy (IMRT) represents one of the most significant technical advances in radiotherapy since the advent of the medical linear accelerator, as it enables the clinical application of highly conformal and non-convex dose distributions. However, these advances do not come without risk, as IMRT represents a new paradigm in the radiation therapy process, particularly in terms of treatment quality assurance. This work focuses on one of the quality assurance aspects of IMRT, related to quality controls of multileaf collimators. A set of tests is presented to routinely evaluate the performance of this component of the accelerator to ensure the reliability and safety of the treatments. The results obtained show that most of the parameters studied comply with internationally accepted tolerances, so it is concluded that this component is suitable for use in IMRT treatments. (author)

Availability note (English)

Also available on line at the web http://nucleus.cubaenergia.cu/index.php/nucleus/article/view/756/1010

Additional details

Additional titles

Original title (Spanish)
Controles de calidad MLC específico para IMRT

Publishing Information

Journal Title
Nucleus (Havana)
Journal Issue
no.71
Journal Page Range
p. 10-15
ISSN
0864-084X

Optional Information

Notes
3 figs., 4 tabs., 9 refs.