A study of a reduction of patient's radiation exposure by using the new ortho screen film system. 3. A study of Kodak InSight EF/RA imaging system

  • 1. Tokushima Univ. (Japan). School of Medical Science

Description

The developing temperature dependency of Kodak InSight EF/RA Image System (IEF system) was evaluated by measuring characteristic curve, resolution properties and noise RMS. Furthermore, the possibility of the reduction of the patient's exposure without decrease in image quality was evaluated. The physical imaging properties of screen/film systems were affected by the developing temperature. Except the gross fog, other imaging properties of IEF system on developing temperature was small. Especially it was confirmed that the developing temperature dependency of granularity of IEF system was less than those of the other new ortho screen film systems (Konica ES-C medical x-ray film of EX system and Fuji UR-1 medical x-ray film of AD system). Therefore, there is the possibility of the reduction of the patient's exposure by using the higher developing temperature than the traditional situation after due consideration of increasing the gross fog. (author)