Published August 2018 | Version v1
Journal article

A study on the performance evaluation of portable radiation shielding apparatus

  • 1. Dept. of Radiology, Yeongdong Hospital, Seoul (Korea, Republic of)
  • 2. Dept. of Radiological Science, Hanseo University, Seosan (Korea, Republic of)

Description

When using a mobile X-ray unit, primary radiation creates medical images and secondary radiation scatters in many directions, which reduces image quality and causes exposure to patients, care givers and medical personnel. The purpose of this study was to develop a radiation shielding system for effectively shielding secondary radiation and evaluate its effectiveness. Using a mobile X-ray unit, spatial dose according to presence of human equivalent phantom and spatial dose using the developed shielding device were measured, and the phantom at 80 cm equidistance from center of X-ray was compared with spatial dose according to use of a shield. Measurements were taken at intervals of 10 cm every 30° from the head direction(-90°) to the body direction(+90°). In the spatial dose measurement with and without the phantom, when the human equivalent Phantom was used, the spatial dose was increased by 40% in all directions from 40 cm to 100 cm from the central X-ray, and about 88% of the space dose was reduced when using the developed shields with the phantom. The equidistance dose at 80 cm from the central X-ray was increased by 39% from 5.1±0.26 μ Gy to 7.1±0.15 μGy when the human equivalent phantom was used, and when phantom was used and shielding was used, the spatial dose was reduced by about 90% from 7.1±0.15 μGy to 0.7±0.07 μGy. The spatial dose of natural radiation was measured to be about 0.2±0.04 μGy when using the developed shielding with Phantom at a distance of 1 m or more. It is expected that by using the developed shielding system, it will be possible to effectively reduce secondary radiation dose received in all directions and to ensure safe imaging

Additional details

Publishing Information

Journal Title
Journal of Radiological Science and Technology
Journal Volume
41
Journal Issue
4
Series
21 refs, 5 figs, 1 tab
Journal Page Range
p. 289-295
ISSN
2288-3509

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
50047785
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
EFFICIENCY; EQUIPMENT; IMAGES; PATIENTS; PERFORMANCE; PHANTOMS; RADIATION DOSES; RADIATIONS; SHIELDING
Descriptors DEC
DOSES; MOCKUP; STRUCTURAL MODELS