Published September 1988 | Version v1
Journal article

ECG-triggered MR blood flow velocity measurement in the carotid arteries: Comparison with duplex sonography

  • 1. Siemens A.G., Erlangen (Germany, F.R.). Unternehmensbereich Medizinische Technik
  • 2. Klinikum Grosshadern, Muenchen (Germany, F.R.). Radiologische Klinik und Poliklinik

Description

Blood flow velocity in the common carotid arteries were measured for 30 vessels by MRI with a multiecho SE-sequence and compared with the results from duplex ultrasound. MR blood flow velocity was measured for a time interval of 390 ms every 30 ms using a ECG-triggered data acquisition. The time-velocity curves measured by MRI and duplex ultrasound were in good agreement. The correlation coefficient for the peak velocity was r = 0.83. Mean peak velocity as measured by MRI was 59±8% of peak velocity as measured by duplex ultrasound. (orig.)

Abstract (German)

Die Blutflussgeschwindigkeit in der A. carotis communis wurde in 30 Gefaessen EKG-getriggert in einem zeitlichen Raster von jeweils 30 ms ueber ein Zeitintervall von 390 ms mit Hilfe einer Multiecho-SE-Sequenz kernspintomographisch gemessen und mit den Ergebnissen der Duplex-Sonographie verglichen. Beide Untersuchungstechniken fuehrten zu vergleichbaren zeitlichen Geschwindigkeitsverlaeufen. Der Korrelationskoeffizient fuer die mit beiden Verfahren gemessenen Maximalgeschwindigkeiten betrug r = 0,83. Die mit der Kernspintomographie gemessene mittlere Stroemungsgeschwindigkeit betrug 59±8% der mit der Duplex-Sonographie gemessenen Spitzengeschwindigkeit. (orig.)

Additional details

Additional titles

Original title (German)
EKG-getriggerte kernspintomographische Blutflussgeschwindigkeitsmessung in den Karotiden: Vergleich mit Duplex-Sonographie