The stabilized Michelson interferometer was developed in order to measure micro dynamic displacement at the surface of solids due to elastic wave propagation. The stabilizer was designed to compensate light path disturbances using a reference mirror driven by piezoelectric actuator. Using stabilizer, the effect of external vibration was reduced and the quadrature condition was satisfied. As the results, the output of photodetector had maximum sensitivity and linearity. The minimum detectable displacement was 0.3nm at the band width of 10 MHz. The epicentral displacements due to the glass capillary breaks and the steel ball drop impact were measured using the developed interferometer and the results were compared with the calculated one

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Nondestructive Testing
Journal Volume
13
Journal Issue
4
Series
9 refs, 11 figs
Journal Page Range
p. 32-41
ISSN
1225-7842

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
42102708
Subject category
S42: ENGINEERING;
Descriptors DEI
COMPUTER CALCULATIONS; DESIGN; DETECTION; ELASTICITY; INTERFEROMETERS
Descriptors DEC
MEASURING INSTRUMENTS; MECHANICAL PROPERTIES