Published January 1994
| Version v1
Journal article
Detection of Elastic Waves Using Stabilized Michelson Interferometer
Creators
- 1. Korea Research Institute of Standards and Science, Daejeon (Korea, Republic of)
- 2. Dong Sin University, Gwangju (Korea, Republic of)
Description
The stabilized Michelson interferometer was developed in order to measure micro dynamic displacement at the surface of solids due to elastic wave propagation. The stabilizer was designed to compensate light path disturbances using a reference mirror driven by piezoelectric actuator. Using stabilizer, the effect of external vibration was reduced and the quadrature condition was satisfied. As the results, the output of photodetector had maximum sensitivity and linearity. The minimum detectable displacement was 0.3nm at the band width of 10 MHz. The epicentral displacements due to the glass capillary breaks and the steel ball drop impact were measured using the developed interferometer and the results were compared with the calculated one
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society for Nondestructive Testing
- Journal Volume
- 13
- Journal Issue
- 4
- Series
- 9 refs, 11 figs
- Journal Page Range
- p. 32-41
- ISSN
- 1225-7842
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 42102708
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- COMPUTER CALCULATIONS; DESIGN; DETECTION; ELASTICITY; INTERFEROMETERS
- Descriptors DEC
- MEASURING INSTRUMENTS; MECHANICAL PROPERTIES