Published September 11, 2007 | Version v1
Journal article

A high-efficiency Transition Radiation Detector for high-counting-rate environments

  • 1. National Institute for Nuclear Physics and Engineering, Bucharest (Romania)
  • 2. University of Muenster (Germany)
  • 3. Gesellschaft fuer Schwerionenforschung, Darmstadt (Germany)

Description

A prototype Transition Radiation Detector (TRD) with a new configuration was built and tested. The prototype consists of two individual multiwire proportional chambers (MWPC) that share a thin common central pad readout electrode. Measurements with a 55Fe source and e, π and p of 1.5 GeV/c showed a very good energy, position resolution and a better e/π discrimination compared to the standard structure with a single MWPC. No significant deterioration of the resolutions is observed up to counting rates of 2x105particlescm-2s-1. These results open the possibility of constructing TRDs with a high e/π discrimination and granularity even for high-counting rate experiments with a reasonable number of layers

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2007.06.013

Additional details

Identifiers

DOI
10.1016/j.nima.2007.06.013;
PII
S0168-9002(07)01217-X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
579
Journal Issue
3
Journal Page Range
p. 961-966
ISSN
0168-9002
CODEN
NIMAER