Published September 1988 | Version v1
Journal article

An evaluation of the CRNL/CFFTP HT/HTO discriminating monitor

  • 1. Harwell Lab., Oxon (UK)

Description

A prototype discriminating HT/HTO monitor was loaned to Harwell by CRNL to evaluate for possible use at the JET facility in the UK. The instrument was exposed to pure HT and HTO inputs at various concentrations, and it readings compared with those of a non-discriminating Harwell Ion Chamber connected in series at the input. The rate of response, accuracy, separation factors and possible memory effects were studied at various humidities at ambient temperature, and the effects of varying proceses- and sample-air rates were investigated. Generally, the instrument was found to work well, although the response times for the HTO channel were generally rather slow (20-30 min to 90% of its asymptotic value), so that central sampling of a number of points in sequence via sampling lines and a manifold would be slow

Additional details

Publishing Information

Journal Title
Fusion Technology
Journal Volume
14
Journal Issue
2
Series
Fusion Technol.
Journal Page Range
1054-1057
ISSN
0748-1896