Published October 1998 | Version v1
Journal article

Accuracy of cortical and trabecular bone measurements with peripheral quantitative computed tomography (pQCT)

  • 1. Osteoporosis and Arthritis Research Group, Department of Radiology, University of California - San Francisco, San Francisco, CA (United States)
  • 2. Hand and Microvascular Research Laboratory, Department of Orthopedic Surgery, University of California - San Francisco, San Francisco, CA (United States)

Description

In order to assess the accuracy of peripheral QCT (Stratec XCT 960) we analysed scans of the European Forearm Phantom and another phantom consisting of K2HPO4 encased in aluminium tubes to simulate cortical walls. Additionally 14 cadaveric forearm specimen scans were compared to CT scans acquired on a GE9800Q. The accuracy for density assessment of the European Forearm Phantom was better than 3%. A small increase in density was observed with increasing thickness of the aluminium wall (10% for each mm). Density measurements within the wall were confounded by limited spatial resolution. For a thickness of less than 4 mm, the density within the wall was underestimated by up to 40%. The measurement of mineral content was not influenced by this effect and showed an accuracy error of less than 6%. The agreement of density measurements on the different CT systems was very strong (R2>0.96; RMSE<6.2%). Our findings suggest that the Stratec pQCT scanner very accurately measures volumetric trabecular and total bone mineral densities at the distal radius while the assessment of cortical density is associated with considerable inaccuracies due to limited spatial resolution. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
43
Journal Issue
10
Journal Page Range
p. 2873-2883
ISSN
1361-6560

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43051368
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; BONE TISSUES; COMPUTERIZED TOMOGRAPHY; DENSITY; DIAGNOSTIC USES; MINERALS; PHANTOMS; SKELETON
Descriptors DEC
ANIMAL TISSUES; BODY; CONNECTIVE TISSUE; DIAGNOSTIC TECHNIQUES; MOCKUP; ORGANS; PHYSICAL PROPERTIES; STRUCTURAL MODELS; TOMOGRAPHY; USES

Optional Information

Notes
30 refs; This record replaces 31034735