A study on PAE of tube voltage of diagnostic x-ray units
Creators
- 1. An San Junior College, Ansan (Korea, Republic of)
- 2. Sin Gu Junior College, Sungnam (Korea, Republic of)
Description
In this paper, the PAE of tube voltage of 93 diagnostic X -ray units those are using at the 19 general hospitals and 9 local clinics in the area of Sungnam, Inchon, and Seoul have been surveyed. The results are summarized as following. 1. PAE of tube Voltage 70, 100KV at the 56 (60%) X-ray units were ±7% below, at the 8 units (9% ), were ±7 % above, and those 29 X-ray units (31%) were ±7 % below or above with the 70, 100KV tube Voltage in a unit. 2. PAE ±7% below were 68 X-ray units, at the low tube voltage (70KV), and 76 X-ray units at the high tube Voltage (100KV). 3. For six year-use abofe, used X-ray units were PAE ±7% above, and so, the longer used years the more X - ray units became declined in those performance. 4. PAE of tube Voltage were better efficiency at the large capacity 30 X-ray units than small capacity 10. 5. PAE ±7 % above in were only 8 units among 84 diagnostic X-ray units in the general hospitals
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society of Radiological Technology
- Journal Volume
- 16
- Journal Issue
- 2
- Series
- 10 refs, 4 figs, 3 tabs
- Journal Page Range
- p. 45-50
- ISSN
- 1226-2854
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 40075214
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DIAGNOSIS; ELECTRIC POTENTIAL; HOSPITALS; PERFORMANCE; SURVEY MONITORS; TUBES
- Descriptors DEC
- BUILDINGS; MEASURING INSTRUMENTS; MEDICAL ESTABLISHMENTS; MONITORS; RADIATION MONITORS